Please login to use this feature. You can use this feature to add the product to your favourite list.
Close
You have added this product to your favorite list. Check My Favourite
Close
You have removed this product from your favourite list.
Close
Please login to use this feature. You can use this feature to add the company to your favourites list.
Close
This company has been added successfully. Check My Favourite
Close
This company has been removed from your favourite list.
Close
Please login to use this feature. You can use this feature to add the company to your inquiry cart.
Close
This company has been added to your inquiry cart.
Close
This company has been removed from your inquiry cart.
Close
This product has been added to your inquiry cart.
Close
This product has been removed from your inquiry cart.
Close
Maximum number of Product/Company has been reached in inquiry cart.
Close
Obsnap Calibration Sdn Bhd
Obsnap Calibration Sdn Bhd 201701008420 (1222585-W)
SSM
Location:

Selangor Subang Jaya

Business Nature:

Services

Calibration of humidity chamber - Obsnap Calibration Sdn Bhd

Calibration of humidity chamber

11-Jan-2025

Calibration of a humidity chamber involves verifying and adjusting its performance to ensure it produces accurate and consistent humidity levels. Here's a step-by-step guide:


1. Preparation

  1. Read the Manual: Familiarize yourself with the chamber's operation and calibration instructions provided by the manufacturer.
  2. Gather Required Tools:
    • Hygrometer or reference instrument (calibrated and traceable to national standards).
    • Data logger (optional for continuous monitoring).
    • Calibration salts or a humidity generator (if required).
    • Tools for making adjustments, as per the chamber's specifications.
  3. Stabilize the Environment: Ensure the chamber is in a stable ambient environment free from significant temperature or humidity fluctuations.

2. Perform Calibration

Option A: Using Reference Instrument

  1. Place the Reference Instrument:

    • Position the calibrated hygrometer inside the chamber at the test point.
    • If the chamber is large, place sensors in multiple locations to ensure uniformity.
  2. Set the Humidity:

    • Program the chamber to specific humidity levels (e.g., 30%, 50%, 70%, and 90% relative humidity).
    • Let the chamber stabilize at each set point.
  3. Record the Readings:

    • Compare the chamber's displayed humidity with the reference instrument's reading.
    • Note deviations at each set point.
  4. Repeat for Temperature Variation:

    • If the chamber has temperature control, perform the same tests at different temperatures to verify combined accuracy.

Option B: Using Saturated Salt Solutions

  1. Prepare Salt Solutions:

    • Use salts like sodium chloride (NaCl) or magnesium chloride (MgCl ) to create known humidity levels in sealed containers.
    • Refer to a salt reference table for expected RH levels at specific temperatures.
  2. Place Solutions Inside:

    • Place the solution container in the chamber and allow stabilization.
  3. Measure Humidity:

    • Compare the chamber's readings with the expected humidity from the salt solution.

3. Adjust the Chamber (if needed)

  • Use the calibration interface or manual adjustments to correct deviations.
  • Follow the manufacturer’s procedure for entering calibration offsets.

4. Verification

  • After adjustment, recheck all set points to confirm the accuracy.
  • Perform a stability test by running the chamber for extended periods and observing deviations over time.

5. Document Calibration

  • Record:
    • The reference instrument details (serial number, last calibration date).
    • The observed deviations and corrections made.
    • Environmental conditions during calibration.
    • The date and technician's name.

6. Schedule Routine Calibration

  • Regularly calibrate the chamber (e.g., annually or as specified by quality standards).

By maintaining proper calibration, your humidity chamber will provide reliable and precise conditions for testing or storage.

Main Office

Obsnap Calibration Sdn Bhd 201701008420 (1222585-W)
No. A-7-3 Floor, Jalan Serai Wangi F 16/F, Alam Avenue, Seksyen 16, 40200 Shah Alam, Selangor, Malaysia.

Tel:

Email:
Website: https://www.obsnapcalibration.com
Website: https://obsnapcalibration.newpages.com.my/
Website: https://obsnapcalibration.onesync.my/

Other Office

Melaka Branch
No 49-1, Jalan KNMP2A, Kompleks Niaga Melaka Perdana, 75450 Bukit Katil, Melaka.

Tel:

Browse by : Home - Classifieds - Companies - Location - Tags - Products - News & Promotion - Job Vacancy - Mobile Website - Google - SEO Results

NEWPAGES

  • BR 22118
  • US 21942
  • GB 15102
  • CA 8730
  • AU 8074
  • IE 5610
  • TR 3387
  • NZ 1910
People Online
Seni Jaya Logo
Brochure
Download
Our PackageContact Us