Reliable stability
- Monitored parameters include: Expiratory tidal volume (Vte), Respiratory rate(Res. Rate),Minute ventilation (Min Vent), Leakage (Leak)
- Quick and easy settings when in use, including: IPAP, EPAP, Res.Rate, E/I Rate,I sens, E sens, Rise time.
- Innovative tracking technology ensures accurate and comfortable therapy.
- Unique sensitivity setting offers individual therapy algorithm, makes every userget the most comfort.
- Auto on/Auto off.
- Alert when accidental power off.
- In time alert function when mask/tubing off line.
- Automatic leakage and altitude compensation.
Ergonomic design
- Integrated and knock-down designed InH2 heated humidifier.
- DC 24V powered and infrared controlled humidifier make user safe andcomfortable.
- Patented anti-countercurrent water tank.
- Unique Delay-off feature protects RESmart BPAP from humidity hazard.
- Powerful embedded flash memory and SD card stores more than 30 nights of fullraw airflow data and 365 nights of user blog.
Techical Info
Device Size
| Dimensions: |
220 × 194 × 112 mm
|
|
313 × 194 × 112 mm (with InH2 heated humidifier)
|
|
| Weight: |
< 2.2 kg
|
|
< 3 kg (with InH2 heated humidifier)
|
Product Use, Transport and Storage
| Operation |
Transport and Storage
|
|
| Temperature: | 5 to 35 °C |
-20 to 55 °C
|
| Humidity: | ≤80% Non-condensing |
≤93% Non-condensing
|
| Atmospheric Pressure: | 860 to 1060 hPa |
500 to 1060 hPa
|
Model Comparison
| Product Model | Pressure |
Work Mode
|
| RESmart BPAP 20S | 4-20 cm H2O |
CPAP, S
|
| RESmart BPAP 20A | 4-20 cm H2O |
CPAP, S, Auto S
|
| RESmart BPAP 20T | 4-20 cm H2O |
CPAP, S, T, S/T
|
| RESmart BPAP 25 | 4-25 cm H2O |
CPAP, S
|
| RESmart BPAP 25A | 4-25 cm H2O |
CPAP, S, Auto S
|
| RESmart BPAP 25T | 4-25 cm H2O |
CPAP, S, T, S/T
|
| RESmart BPAP 30T | 4-30 cm H2O |
CPAP, S, T, S/T
|
SD card capacity
Pressure Range
Pressure Stability
| Test Pressure (cm H2O) | 4 | 8 | 12 | 16 | 20 |
| Average flow at the patient connection port (L/min) For 25/ 25A/ 25T: | 70.1 | 72.1 | 72.3 | 70.1 |
73.4
|
| Test Pressure (cm H2O) | 4 | 10 | 15 | 20 | 25 |
| Average flow at the patient connection port (L/min) For 30T: | 70.1 | 73.0 | 64.0 | 73.4 |
74.6
|
| Test Pressure (cm H2O) | 4 | 10.5 | 17 | 23.5 |
30
|
| Average flow at the patient connection port (L/min) | 70.1 | 73.0 | 70.5 | 71.7 | 72.8 |