Please login to use this feature. You can use this feature to add the product to your favourite list.
Close
You have added this product to your favorite list. Check My Favourite
Close
You have removed this product from your favourite list.
Close
Please login to use this feature. You can use this feature to add the company to your favourites list.
Close
This company has been added successfully. Check My Favourite
Close
This company has been removed from your favourite list.
Close
Please login to use this feature. You can use this feature to add the company to your inquiry cart.
Close
This company has been added to your inquiry cart.
Close
This company has been removed from your inquiry cart.
Close
This product has been added to your inquiry cart.
Close
This product has been removed from your inquiry cart.
Close
Maximum number of Product/Company has been reached in inquiry cart.
Close
MSP Metrology (M) Sdn Bhd
MSP Metrology (M) Sdn Bhd 201301013617 (1043455-P)
Business
Hours
Monday - Friday 8:30 AM - 6:00 PM
Saturday - Sunday Closed
We′re closed on Public Holiday
Open Closed
Onesync Platinum SSM
60167220677
+60167220677 WeChat QR
WeChat QR

Wechat ID: +60167220677

Introduction to Portable Leeb Hardness Tester

22-Jun-2023

The portable leeb hardness tester, commonly referred to as the rebound method, determines hardness from voltages generated by the electromagnetic induction principle. Pressing the button on top of the impact device releases a spring that propels an impact body through a guide tube toward the part’s surface, as shown in Fig. 1.

The impact body passes through a coil surrounding the guide tube. The motion of the impact body’s magnet through the coil creates a voltage across the coil windings. When the impact body’s indenter strikes the part, a portion of its energy is lost in the material. The softer the material, the greater the energy loss and the slower the rebound velocity.

The magnet returning through the coil creates a second voltage whose amplitude is relative to the velocity of the impact body. Figure 2 illustrates the voltage generated by the impact body traveling through one coil.

 

Fig. 1 — Illustration showing how the impact body travels through the guide tube of a rebound method portable hardness testing machine.
 

A variety of impact devices are available, with the D device being the most common model. The D device has a 3-mm-diameter tungsten carbide ball indenter and a spring force to create an indentation size suitable for homogeneous materials. The E device provides the same spring force as the D device but uses a 3-mm-diameter diamond indenter for testing parts in excess of 50 HRC. The G device uses a larger 5-mm-diameter tungsten carbide ball indenter and a heavier spring to create a larger indentation, which is necessary for testing coarse-grained forgings and castings. Some manufacturers provide variants of the D device for use when access is limited.
 

Rebound Applications

The rebound method requires large, massive parts. Due to its large indentation size, it is the best solution for coarse-grained castings and forgings. Examples of applications include the following:

  • Steel mill rolls
  • Engine blocks
  • Shafts or axles
  • Extrusion dies
  • Crane wheel pulleys.
 

Factors Influencing the Rebound Method

  • The rebound method creates a large force of short duration during impact. Thin and lightweight parts may yield, causing erroneous and random readings. For example, the D device requires a minimum weight of 5 kg and a minimum wall thickness of 20 mm. Smaller parts can be tested but require the aid of an additional support fixture.
  • Gravity affects the velocity of the impact body; therefore, as the test direction changes, a correction factor is required. Most testers require the operator to manually input any changes to the direction prior to performing a test. However, more advanced instruments sense the direction and compensate automatically.
  • The measured HL value is not frequently specified in test procedures. To display the common Brinell and Rockwell scales, the operator must choose the appropriate material group to select a conversion table created for a specific Young’s modulus. Since conversion tables are based on empirical relationships, they must be considered only approximations.
  • The tungsten carbide ball indenter of the D and G devices may become flattened, causing high readings, or, for all models, the guide tube may require cleaning to prevent low readings. Periodic checks on a test block will detect these conditions.

 

Above sharing just for your reference. When you’re interested in investing a hardness tester to improve your quality process, select a supplier who offer a quality service and meet your specific needs.

If you have any enquiry with hardness tester, welcome to contact us by email: enquiry@mspmetrology.com

Main Office

MSP Metrology (M) Sdn Bhd 201301013617 (1043455-P)
10, Jalan Gemilang 9, Taman Perindustrian Cemerlang, 81800 Ulu Tiram, Johor, Malaysia.

Tel:

Email:
Website: https://www.mspmetrology.com.my
Website: https://mspmetrology.newpages.com.my/
Website: https://mspmetrology.onesync.my/

Other Office

KL/Shah Alam Branch
18-23A, Plaza Azalea, 6, Persiaran Bandaraya , Seksyen 14, 40000 Shah Alam, Selangor, Malaysia.

Tel:

Union Technologies Pte Ltd
71 Bukit Batok Crescent #07-11/12, Prestige Centre, Singapore 658071.

Tel:
Fax:

Penang Branch
35-5-6, Golden Triangle 2 Commercial Square, Jalan Dato Ismail Hashim, 11900 Bayan Lepas, Penang, Malaysia.

Tel:

Browse by : Home - Classifieds - Companies - Location - Tags - Products - News & Promotion - Job Vacancy - Mobile Website - Google - SEO Results

NEWPAGES

  • US 21119
  • BR 17077
  • GB 13189
  • CA 10147
  • AU 9238
  • IE 5680
  • IN 3351
  • NZ 2214
People Online
Seni Jaya Logo
Brochure
Download
Our PackageContact Us