Features
- Broad measuring range of -195 to +1350 °C possible thanks to connectable probes (optional)
- WiFI data logger system with easy installation and operation
- Call up measurement data at any time via Smartphone, tablet or PC
- Limit value violation alarms by e-mail or (optionally) SMS
Temperature - TC Type K (NiCr-Ni)
Measuring range | -195 to +1350 °C |
---|---|
Accuracy | ±(0.5 + 0.5 % of mv) °C |
Resolution | 0.1 °C |
Temperature - TC Type T (Cu-CuNi)
Measuring range | -200 to +400 °C |
---|---|
Accuracy | ±(0.5 + 0.5 % of mv) °C |
Resolution | 0.1 °C |
Temperature - TC Type J (Fe-CuNi)
Measuring range | -100 to +750 °C |
---|---|
Accuracy | ±(0.5 + 0.5 % of mv) °C |
Resolution | 0.1 °C |
General technical data
Weight | 240 |
---|---|
Dimensions | 95 x 75 x 30.5 mm |
Operating temperature | -30 to +50 °C |
Protection class | IP54 |
Measuring rate | Advanced: 1 min … 24 h |
Door contact | - |
Communication rate | Advanced: 1 min … 24 h |
WLAN Connectivity | Signal transmission: wireless, frequency band: 2.4 GHz, supported wireless LAN standards: IEEE 802.11 b/g/n and IEEE 802.1X; Possible encryption methods: without encryption, WEP, WPA, WPA2, WPA2 Enterprise - The data loggers communicate via the standard protocol MQTT and are SNTP time synchronization-capable. |
Connection external | External temperature probes |
Power supply | 4 x AA AlMn batteries; mains unit optional; for temperatures below -10 °C please use Energizer batteries 0515 0572 |
Memory | 10,000 measurement values / channel |
Battery life | 12 months (typical value, depending on the wireless LAN infrastructure) at +25 °C, 15-minute measuring cycle and standard communication cycle at -30 °C, 15-minute measuring cycle and standard communication cycle with Energizer batteries 0515 0572 |
Storage temperature | -40 to +70 °C (without batteries) |