¡¾DIRECTIONS FOR USE¡¿
Allow test dipstick, specimen, and/or controls to equilibrate to room temperature (15-30°C) prior to testing.
1. Bring the pouch to room temperature before opening it. Remove the test dipstick from the sealed pouch and use it as soon as possible. Best results will be obtained if the assay is performed within one hour.
2. For Serum or Plasma specimen: Hold the dropper vertically and transfer 1 drop of serum or plasma (approximately 25 μl ) to the specimen area, then add 2 drops of buffer (approximately 120 μl ),and start the timer, see illustration below.
For Venipuncture Whole Blood specimen: Hold the dropper vertically and transfer 2 drops of whole blood (approximately 50 μl ) to the specimen area, then add 2 drops of buffer (approximately 120 μl ), and start the timer. See illustration below.
For Fingerstick Whole Blood specimen:
· To use a capillary tube: Fill the capillary tube and transfer approximately 50 μl of fingerstick whole blood specimen to the specimen area of test dipstick, then add 2 drops of buffer (approximately 120 μl ) and start the timer. See illustration below.
· To use hanging drops: Allow 2 hanging drops of fingerstick whole blood specimen (approximately 50 μl ) to fall into the specimen area of test dipstick, then add 2 drop of buffer (approximately 120 μl ) and start the timer. See illustration below.
3. Wait for the colored line(s) to appear. The test result should be read at 10 minutes.
4. Do not interpret the result after 20 minutes.

|