¡¾DIRECTIONS FOR USE¡¿
Allow the test, specimen, buffer and/or controls to reach room temperature (15-30°C) prior to testing.
1. Bring the pouch to room temperature before opening it. Remove the test cassette from the sealed pouch and use it as soon as possible. Best results will be obtained if the assay is performed within one hour.
2. Place the cassette on a clean and level surface. Hold the dropper vertically, draw the specimen (whole blood/ serum/plasma) up to the point as shown in illustration below (approximately 10μl). Transfer the specimen to the sample well (S) each, then hold the buffer bottle vertically and add 2 drops of buffer (approximately 80μl) to the sample well (S), and start the timer. See the illustration below.
3. Wait for the colored line(s) to appear. The result should be read at 10 minutes. Do not interpret results after 20 minutes.

| 
					 Cat. No.  | 
				
					 Product Description  | 
				
					 Specimen  | 
				
					 Format  | 
				
					 Kit Size  | 
				
					 Cut-Off  | 
				
					 Status  | 
			
| 
					 IMPM-402  | 
				
					 Mycoplasma Pneumoniae IgM Rapid Test Cassette  | 
				
					 WB/S/P  | 
				
					 Cassette  | 
				
					 40T  | 
				
					 See Insert  | 
				
					 CE  | 
			


            
            
        
        
        
            
		