¡¾DIRECTIONS FOR USE¡¿
Allow the test, specimen, buffer and/or controls to reach room temperature (15- 30°C) prior to testing.
1. Bring the pouch to room temperature before opening it. Remove the test cassette from the sealed pouch and use it as soon as possible.
2. Place the cassette on a clean and level surface.
For Serum or Plasma specimen:
· Hold the dropper vertically and transfer 1 drop of serum or plasma (approximately 25 μL) to the specimen well of test cassette, then add 3 drops of buffer (approximately 120ul), and start the timer. See illustration below.
For Venipuncture Whole Blood specimen:
· Hold the dropper vertically and transfer 1 drop of whole blood (approximately 25 μL) to the specimen well, then add 3 drops of buffer (approximately120 ul), and start the timer. See illustration below.
3. Wait for the colored line(s) to appear. Read results at 10 minutes. Do not interpret the result after 20 minutes.

| 
				 Cat. No.  | 
			
				 Product Description  | 
			
				 Specimen  | 
			
				 Format  | 
			
				 Kit Size  | 
			
				 Cut-Off  | 
			
				 Status  | 
		
| 
				 ITE-402  | 
			
				 Tetanus Rapid Test Cassette  | 
			
				 WB/S/P  | 
			
				 Cassette  | 
			
				 40 T  | 
			
				 See Insert  | 
			
				 CE  | 
		


            
            
        
        
        
            
		
BR                                14210                            
VN                                9671                            
US                                3922                            
AR                                2318                            
CN                                1536                            
MY                                1156                            
EC                                939                            
SG                                763                            
        
        