Battery powered datalogger with built-in atmospheric pressure sensor and 4 universal inputs for measurement of temperature, humidity, 0-10V voltage, 0-20mA current, two-state signal and for counting pulses. Multilogger can be installed permanently to Ethernet network or work as portable device.
Included is traceable calibration certificate in accordance with EN ISO/IEC17025.
The Multilogger is a battery operated datalogger capable of measuring and recording sensor and electrical values through its inputs. Datalogger can be connected to Ethernet. Then measured values can be viewed on a web browser, as well as stored and downloaded for later analysis.
INCLUDED IN DELIVERY:
EXTERNAL PROBES ARE NOT INCLUDED IN DELIVERY
| TEMPERATURE SENSOR - external probe Pt1000 | |
| Measuring range | -200 to +600 °C | 
| Accuracy | 
				±0.2 °C (-200 to +100 °C); <br>±0.2 % from reading from +100 to +600 °C  | 
		
| Resolution | 0.1 °C | 
| 
				TEMPERATURE+HUMIDITY SENSOR - external probe T+RH DIGI  | 
			|
| Measuring range | according to the connected probe Digi/M | 
| Accuracy | according to the connected probe Digi/M | 
| Resolution | 0.1 °C | 
| DEW POINT | |
| Measuring range | according to the connected probe Digi/M | 
| Accuracy | according to the connected probe Digi/M | 
| Resolution | 0.1 °C | 
| ATMOSPHERIC PRESSURE SENSOR | |
| Measuring range | 600 hPa to 1100 hPa | 
| Accuracy | ±1,3 hPa at 23 °C | 
| Resolution | 0.1 hPa | 
| DC VOLTAGE | |
| Measuring range | 0 to 10 V | 
| Accuracy | ±10 mV | 
| Resolution | better than 100 µV | 
| DC CURRENT | |
| Measuring range | 0 to 20 mA | 
| Accuracy | ±20 µA | 
| Resolution | better than 0,1 µA | 
| BINARY SIGNAL | |
| Input voltage for level “switched on”/ | <0,4 V/> 2 V | 
| Minimum/maximum input voltage | 0 V/+30 V DC | 
| I mpedance of the “switched on”/ | < 10 kOhm/> 300 kOhm | 
| COUNTER INPUT | |
| Range of counter | 
				24 bit (16 777 215) the counter overflow can be allowed  | 
		
| Maximum input frequency | 200 Hz | 
| GENERAL TECHNICAL DATA | |
| Operating temperature | -10 to +60 °C | 
| Channels | 
				internal atmospheric pressure sensor, 2x external temperature probe or connectable temperature+humidity probe, <br>2x binary, counter, voltage and current input  | 
		
| Counted values | 
				dew point, absolute humidity, specific humidity, mixing ratio, specific enthalpy  | 
		
| Memory | 
				1 000 000 values (noncyclic record), 600 000 values (cyclic record)  | 
		
| Recording interval | 1s to 24 hours | 
| Display and alarm refresh | each 10 s | 
| Recording mode | 
				noncyclic - data logging stops after filling the memory<br>cyclic - after filling memory oldest data is overwritten by new  | 
		
| Available temperature units | degrees Celsius, degrees Fahrenheit | 
| Communication interface | Ethernet, USB, RS232 | 
| Ethernet communication protocols | WWW, SNMPv1, XML, DATALINK | 
| Alarm protocols | |
| Power | 
				AC adapter 230Vac/5Vdc (neccessary for Ethernet interface)  | 
		
| Battery life | 3 - 4 weeks | 
| Protection class | IP20 | 
| Dimensions (without connectors) | 178 x 95 x 37mm, without attached cables | 
| Weight (including batteries) | approx. 380 g | 
| Warranty | 3 years | 


            
            

BR                                17993                            
VN                                14932                            
US                                3435                            
AR                                2973                            
CN                                1532                            
EC                                1126                            
RU                                801                            
BD                                770                            
        
        