HI-2600 pH EC and DO Benchtop meter
HI-2600 benchtop meter enables fast, accurate measurements utilising the Hanna Instruments digital electrode / probes for pH (HI-11310), ORP (HI-36180), conductivity (HI-763100), or dissolved oxygen (HI-764080). Each digital electrode / probe has a unique serial number and the parameter that it measures is automatically identified by the meter.
Once connected, the single sensor or sensors are ready to measure the parameter along with temperature.
The intuitive design simplifies configuration, calibration, measurement, data logging and transfer of data to a USB thumb drive or computer. The benchtop meter offers a Basic Mode that streamlines measurement configuration and is useful for routine applications.
The compact meter (and probe) can be used as a portable device (using its rechargeable battery) or used as laboratory instrumentation.
The meter has a large display with capacitive keys which features a 5.5” (14 cm) LCD. The large display provides a 130 ° wide viewing angle.
Main feature points
- Automatic parameter recognition
- pH/ORP choice of measurement units: pH, mV;
- EC choice of measurement unit: Conductivity: µS/cm, mS/cm
- TDS - ppm (mg/L), g/L
- Salinity - NaCl, PSU, g/L
- DO choice of measurement unit: ppm (mg/L), %
- Adjustable resolution for pH and EC measurements
- CAL Check indicators - HI-11310, HI-763100, HI-764080 digital probes only
- Sensor Check indicators - HI-11311 and HI-12301 digital probes only
- Basic mode for simplified operation
- Dedicated GLP key
- GLP data included with logged data
- Simplified data transfer to a PC
- Up to 8 hour battery life when used as a portable device
- Internal clock and date
Key Features
Digital probe input - simply connect each probe via the 3.5 mm jack, Digital Smart Electrodes are automatically recognised.
Clear, full text readout
- The HI-2600 presents clear, detailed text prompts at the bottom section of the screen.
Two operating modes - HI-2600 can be used in Standard or Basic Operating Modes (pH and EC only). Standard Mode enables all features while Basic Mode displays a simplified screen and features —ideal for routine measurements.
Data logging - The HI-2600 provides ample data storage capacity, allowing you to save up to 1000 log records. Each data set comprehensively includes readings, GLP information, as well as date and time stamps.
GLP - The sensor retains data from the most recent calibration, including date, time, and buffers utilised. When connected to the HI-2600, this GLP information is automatically transferred.
CAL Check - Hanna’s exclusive CAL Check feature analyses the pH electrode’s response in calibration buffers to detect potential issues such as contaminated solutions or dirty electrodes. Upon completing calibration, indicators reflecting the probe’s condition—based on its offset and slope characteristics—are displayed on the measurement screen.
Sensor Check (using HI-11311 and HI-12301 pH probes only) - When paired with compatible Hanna electrodes featuring a matching pin, HI-2600 continuously monitors the impedance of the pH measuring electrode, providing real-time alerts in case of glass breakage. During calibration, Sensor Check displays the junction condition, by evaluating the reference junction resistance.
ORP measurement - Measure ORP (REDOX) in mV units by using a compatible ORP sensor.
Capacitive touch keypad - HI-2600 features sensitive capacitive touch buttons for accurate keystrokes for navigating menus and screens. Since they are part of the screen, the buttons can never get clogged with sample residue.
Easy to read LCD - HI-2600 features a 5.5” (14 cm) LCD display that you can clearly view from over 5 m (16.4’). The large display, with its wide 130° viewing angle, provides one of the easiest to read LCDs in the industry.
Rechargeable battery - The integrated rechargeable battery offers maximum portability with a long 8-hour battery life.
USB-C port for charging and data transfer - The HI-2600 features a streamlined USB-C port that facilitates charging the integrated rechargeable battery, transferring data to your computer, and exporting files directly to a flash drive.
What's included:
- HI-11310 pH digital electrode
- HI-764026 Electrode holder with the following accessories: base plate with integrated pivot pin; cable holder clip, attached
- pH calibration kit: pH 4 buffer solution (2 sachets); pH 7 buffer solution (4 sachets); pH 10 buffer solution sachets (2 sachets)
- Electrode cleaning solution (2 sachets)
- HI-920018 - 1.5m USB-C to USB-C cable
- USB-C power adapter
- Instrument quality certificate
- Probe / electrode quality certificate
- Battery safety insert
- Quick reference guide with QR code for instruction manual download
| Specification | Detail |
| HI2600 | using pH electrode HI11310 |
| pH Range | −2.00 to 16.00 pH −2.000 to 16.000 pH * |
| pH Resolution | 0.01 pH 0.001 pH * |
| pH Accuracy at 25 °C / 77 °F | ±0.01 pH ±0.002 pH * |
| mV Range | ±1000.0 mV |
| mV Resolution | 0.1 mV |
| mV Accuracy at 25 °C / 77 °F | ±0.2 mV |
| Temperature Range (HI11310) | −20.0 to 120.0 °C ** (−4.0 to 248.0 °F) |
| Temperature Resolution (HI11310) | 0.1 °C; (0.1 °F) |
| Temperature Accuracy at 25 °C / 77 °F (HI11310) | ±0.5 °C; (0.9 °F) |
| pH calibration | Automatic, up to three (five*) points Option to select from five (seven*) standard buffers: pH 1.68*, 4.01 or 3.00, 6.86, 7.01, 9.18, 10.01, 12.45* and two custom buffers* |
| HI2600 - using ORP electrode | HI36180 |
| ORP Range | ± 2000.0 mV |
| ORP Resolution | 0.1 mV |
| ORP Accuracy at 25 °C / 77 °F | ±0.2 mV (±999.9 mV) ±1 mV (±2000 mV) |
| Temperature Range (HI36180) | −20.0 to 120.0 °C**(−4.0 to 248.0 °F) |
| Temperature Resolution (HI36180) | 0.1 °C (0.1 °F) |
| Temperature Accuracy at 25 °C / 77 °F (HI36180) | ±0.5 °C (0.9 °F) |
| Calibration | Single point offset |
| Temperature Compensation | Automatic, using integral temperature sensor −5.0 to 100.0 °C (23.0 to 212.0 °F) |
| Logging | Up to 1000* records organized in: • Manual log-on-demand, maximum 200 logs • Manual log-on-stability, maximum 200 logs • Interval logging* |
| HI2600 - using the four-ring conductivity probe | HI763100 |
| EC Range | 0.00 to 29.99 μS/cm 30.0 to 299.9 μS/cm 300 to 2999 μS/cm 3.00 to 29.99 mS/cm 30.0 to 200.0 mS/cm up to 500.0 mS/cm absolute conductivity*** |
| EC Resolution | 0.01 μS/cm 0.1 μS/cm 1 μS/cm 0.01 mS/cm 0.1 mS/cm |
| EC Accuracy at 25 °C / 77 °F | ±1% of reading (±0.05 μS/cm or 1 digit,whichever is greater) |
| TDS Range | 0.00 to 14.99 ppm (mg/L) 15.0 to 149.9 ppm (mg/L)150 to 1499 ppm (mg/L) 1.50 to 14.99 g/L 15.0 to 100.0 g/L up to 400.0 g/L absoluteTDS ‡ (with 0.80 factor) |
| TDS Resolution | 0.01 ppm 0.1 ppm 1 ppm 0.01 g/L |
| TDS Accuracy at 25 °C / 77 °F | ±1% of reading (±0.03 ppm or 1 digit, whichever is greater |
| Salinity* Range | 0.0 to 400.0 % NaCl 2.00 to 42.00 PSU 0.00 to 80.00 g/L |
| Salinity* Resolution | 0.1% NaCl 0.01 PSU 0.01 g/L |
| Salinity* Accuracy at 25 °C / 77 °F | ±1% of reading |
| Temperature Range (HI763100) | −20.0 to 120.0 °C**(−4.0 to 248.0 °F) |
| Temperature Resolution (HI763100) | 0.1 °C (0.1 °F) |
| Temperature Accuracy at 25 °C / 77 °F (HI763100) | ±0.5 °C (0.9 °F) |
| EC calibration | Cell factor calibration Six standards available: 84 µS/cm, 1413 μS/cm 5.00 mS/cm, 12.88 mS/cm, 80.0 mS/cm, 111.8 mS/cm One point offset: 0.00 μS/cm |
| "Salinity calibration* (% NaCl only)" | Single point with HI7037 standard |
| Temperature compensation | • Automatic, −5 to 100 °C (23 to 212 °F) • NoTC, can be selected to measure absolute conductivity |
| Conductivity temperature coefficient* | 0.00 to 6.00 %/°C (EC and TDS only) (1.90 %/°C default) |
| TDS factor* | 0.40 to 0.80 (0.50 default) |
| Logging | Up to 1000* (400) records organized in: • Manual log-on-demand, maximum 200 logs • Manual log-on-stability, maximum 200 logs • Interval logging*, maximum 600 logs (100 lots) |
| Notes | * When operating in Standard Mode only! ‡ Absolute conductivity (TDS) is the conductivity value without temperature compensation. |
| HI2600 - using the Dissolved Oxygen probe | HI764080 |
| DO Range | 0.00 to 45.00 ppm (mg/L) 0.0 to 300.0% |
| DO Resolution | 0.01 ppm (mg/L) 0.1% |
| DO Accuracy at 25 °C / 77 °F | ±1.5% of reading ±1 digit |
| Temperature Range | -20.0 to 120.0 °C * (−4.0 to 24 |
| Temperature Resolution | 0.1 °C (0.1 °F) |
| Temperature Accuracy at 25 °C / 77 °F | ±0.5 °C (0.9 °F) |
| DO calibration | One or two points at 0% (HI7040) and 100% (water-saturated air) |
| DO Pressure compensation | Automatic 450.0 to 850.0 mmHg |
| DO Pressure compensation resolution | 0.1 mmHg |
| DO Salinity compensation | 0 to 40 g/L |
| DO Salinity compensation resolution | 1 g/L |
| Temperature compensation | 0.0 to 50.0 °C (32.0 to 122 °F) |
| Logging | Up to 1000 records organized in: • Manual log-on-demand, maximum 200 logs • Manual log-on-stability, maximum 200 logs • Interval logging**, maximum 600 logs (100 lots) |
| Notes | * Temperature limits will be reduced to actual probe limits. ** When operating in Standard Mode only! |
| PC interface | USB-C |
| Power supply | USB Type C (5 VDC; 500 mA) |
| Environment | 0 to 50 °C (32 to 122 °F) Maximum 95 % RH non-condensing |
| Dimensions | 205 x 160 x 77 mm (8.0 x 6.2 x 3.0 inches) |
| Weight | Approximately 0.85 kg (1.87 lbs.) |