HI-2621 benchtop meter enables fast, accurate pH and ORP measurements utilising the Hanna Instruments digital electrode / probes for pH (HI-11310), ORP (HI-36180). Each digital electrode / probe has a unique serial number and the parameter that it measures is automatically identified by the meter.
Once connected, the single sensor or sensors are ready to measure the parameter along with temperature.
The intuitive design simplifies configuration, calibration, measurement, data logging and transfer of data to a USB thumb drive or computer.
The HI-2621 offers a Basic Mode that streamlines measurement configuration and is useful for routine applications as well as a Standard Operating Mode, where all features and capabilities are enabled.
The compact meter (and probe) can be used as a portable device (using its rechargeable battery) for field measurements or used as laboratory instrumentation.
The meter has a large display with capacitive keys which features a 5.5” (14 cm) LCD. The large display provides a 130 ° wide viewing angle.
Main feature points
- Automatic parameter recognition
- pH/ORP choice of measurement units: pH, mV;
- CAL Check indicators
- Sensor Check indicators
- Basic mode for simplified operation
- Dedicated GLP key
- GLP data included with logged data
- Simplified data transfer to a PC
- Internal clock and date
Digital probe input - simply connect each probe via the 3.5 mm jack, Digital Smart Electrodes are automatically recognised.
Clear, full text readout - The HI-2621 presents clear, detailed text prompts at the bottom section of the screen.
Two operating modes - HI-2621 can be used in Standard or Basic Operating Modes (pH and EC only). Standard Mode enables all features while Basic Mode displays a simplified screen and features —ideal for routine measurements.
Data logging - The HI-2621 provides ample data storage capacity, allowing you to save up to 1000 log records. Each data set comprehensively includes readings, GLP information, as well as date and time stamps.
GLP - The sensor retains data from the most recent calibration, including date, time, and buffers utilised. When connected to the HI-2621, this GLP information is automatically transferred.
CAL Check - Hanna’s exclusive CAL Check feature analyses the pH electrode’s response in calibration buffers to detect potential issues such as contaminated solutions or dirty electrodes. Upon completing calibration, indicators reflecting the probe’s condition—based on its offset and slope characteristics—are displayed on the measurement screen.
Sensor Check (using HI-11311 and HI-12301 pH probes only) - When paired with compatible Hanna electrodes featuring a matching pin, HI-2600 continuously monitors the impedance of the pH measuring electrode, providing real-time alerts in case of glass breakage. During calibration, Sensor Check displays the junction condition, by evaluating the reference junction resistance.
ORP measurement - Measure ORP (REDOX) in mV units by using a compatible ORP sensor.
Capacitive touch keypad - HI-2621 features sensitive capacitive touch buttons for accurate keystrokes for navigating menus and screens. Since they are part of the screen, the buttons can never get clogged with sample residue.
Easy to read LCD - HI-2621 features a 5.5” (14 cm) LCD display that you can clearly view from over 5 m (16.4’). The large display, with its wide 130° viewing angle, provides one of the easiest to read LCDs in the industry.
USB-C port for charging and data transfer - The HI-2621 features a streamlined USB-C port that facilitates transferring data to your computer.
What's included:
- HI-11310 pH digital electrode
- HI-764026 Electrode holder with the following accessories: base plate with integrated pivot pin; cable holder clip, attached
- pH calibration kit: pH 4 buffer solution (2 sachets); pH 7 buffer solution (4 sachets); pH 10 buffer solution sachets (2 sachets)
- Electrode cleaning solution (2 sachets)
- HI-920018 - 1.5m USB-C to USB-C cable
- USB-C power adapter
- Instrument quality certificate
- Probe / electrode quality certificate
- Battery safety insert
- Quick reference guide with QR code for instruction manual download
| Specification | Detail |
| HI2620 and HI2621 | using pH electrode HI11310 |
| pH Range | −2.00 to 16.00 pH −2.000 to 16.000 pH* |
| pH Resolution | 0.01 pH 0.001 pH* |
| pH Accuracy at 25 °C / 77 °F | ±0.01 pH ±0.002 pH* |
| mV Range | ±1000.0 mV |
| mV Resolution | 0.1 mV |
| mV Accuracy at 25 °C / 77 °F | ±0.2 mV |
| Temperature Range (HI11310) | −20.0 to 120.0 °C ** (−4.0 to 248.0 °F) |
| Temperature Resolution (HI11310) | 0.1 °C; (0.1 °F) |
| Temperature Accuracy at 25 °C / 77 °F (HI11310) | ±0.5 °C; (0.9 °F) |
| pH calibration | Automatic, up to three (five*) points Option to select from five (seven*) standard buffers: pH 1.68*, 4.01 or 3.00, 6.86, 7.01, 9.18, 10.01, 12.45* and two custom buffers* |
| HI2620 and HI2621 - using ORP electrode | HI36180 |
| ORP Range | ± 2000.0 mV |
| ORP Resolution | 0.1 mV |
| ORP Accuracy at 25 °C / 77 °F | ±0.2 mV (±999.9 mV) ±1 mV (±2000 mV) |
| Temperature Range (HI36180) | −20.0 to 120.0 °C**(−4.0 to 248.0 °F) |
| Temperature Resolution (HI36180) | 0.1 °C (0.1 °F) |
| Temperature Accuracy at 25 °C / 77 °F (HI36180) | ±0.5 °C (0.9 °F) |
| Calibration | Single point offset |
| Temperature Compensation | Automatic, using integral temperature sensor −5.0 to 100.0 °C (23.0 to 212.0 °F) |
| Logging (HI2621 only) | Up to 1000* records organized in: • Manual log-on-demand, maximum 200 logs • Manual log-on-stability, maximum 200 logs • Interval logging* |
| Notes | *When operating in Standard Mode in HI2621 only! ** Temperature limits will be reduced to actual probe limits. |
| PC interface (HI2621 only) | USB-C |
| Power supply | USB Type C (5 VDC; 500 mA) |
| Environment | 0 to 50 °C (32 to 122 °F) Maximum 95 % RH non-condensing |
| Dimensions | 205 x 160 x 77 mm (8.0 x 6.2 x 3.0 inches) |
| Weight | Approximately 0.85 kg (1.87 lbs.) |