Product Description
The Fluke Calibration P5521 and P5522 Liquid-to-Liquid Separators protect hydraulic deadweight testers and pumps from fluid contamination. A flexible diaphragm separates the calibrator's fluid from the device under test's fluid, allowing calibration in different media.
Key Features & Advantages:
-
Contamination Barrier: A flexible diaphragm physically separates the calibrator's clean fluid from the often-different and potentially corrosive fluid in the device under test.
-
Model-Specific Ranges: The P5521 handles pressures to 10,000 psi (700 bar), while the P5522 is rated for 7,000 psi (480 bar) for use with aggressive fluids like Skydrol.
-
Direct Connection: Connects directly to the test port of hydraulic deadweight testers and comparison test pumps.
-
Calibrate in Working Fluid: Enables devices to be calibrated using their specific operational fluid without risking damage to the primary standard.
Technical Specifications:
-
Brand: Fluke Calibration
-
Model: P5521, P5522
-
Product No.: P5521
-
Max Pressure (P5521): 10,000 psi (700 bar)
-
Max Pressure (P5522): 7,000 psi (480 bar) for aggressive fluids
-
Principle: Diaphragm-based Fluid Separation
Ideal For:
Calibrating hydraulic devices that must operate with specialized or aggressive fluids, protecting valuable calibration equipment from contamination and damage.