Brand : Binder - Germany
Model: FD 60, FD115, FD260
Features
- Temperature range: ambient temperature plus 10 °C to 300 °C
- High temperature accuracy thanks to APT.line™ technology
- Forced convection
- Controller with LCD display
- Electromechanical control of the exhaust air flap
- 2 chrome-plated racks
- Class 2 independent temperature safety device
- Ergonomic handle design
- USB port for recording data
Model | Interior Volume (L) |
FD 56 | 60 |
FD 115 | 116 |
FD 260 | 259 |
Technical Data
Description | FD 56 | FD 115 | FD 260 |
Width net (mm) | 560 | 710 | 810 |
Height net (mm) | 625 | 735 | 965 |
Depth net (mm) | 565 | 605 | 760 |
Measures – Internal dimensions | |||
Width (mm) | 400 | 550 | 650 |
Height (mm) | 440 | 550 | 780 |
Depth (mm) | 345 | 385 | 515 |
Measures | |||
Interior volume (L) | 60 | 116 | 259 |
Net weight of the unit (empty)[kg] | 39 | 54 | 85 |
Permitted load[kg] | 70 | 150 | 270 |
Load per rack[kg] | 30 | 30 | 40 |
Performance Data Temperature | |||
Temperature range 10 °C above ambient temperature to [°C] | 300 | 300 | 300 |
Temperature variation at 150 °C [± K] | 1.7 | 1.7 | 1.9 |
Temperature fluctuation at 150 °C ±K] | 0.3 | 0.3 | 0.5 |
Heating-up time to 150 °C [min] | 15 | 19 | 20 |
Recovery time after 30 seconds door open at 150 °C [min] | 4 | 5 | 6 |
Electrical data | |||
Nominal power (Version 120 V) [kW] | 1.2 | 1.4 | - |
Nominal power (Version 230 V) [kW] | 1.1 | 1.3 | 2.3 |
Nominal power (Version 240 V) [kW] | - | - | 2.5 |
Fixtures | |||
Number of shelves (std./max.) | 2/4 | 2/5 | 2/8 |